Warning: Can't synchronize with repository "(default)" (/home/git/ome.git does not appear to be a Git repository.). Look in the Trac log for more information.
Notice: In order to edit this ticket you need to be either: a Product Owner, The owner or the reporter of the ticket, or, in case of a Task not yet assigned, a team_member"

Task #11707 (new)

Opened 10 years ago

Last modified 10 years ago

BUG: Point ROI Intensity value reads off top right corner — at Initial Version

Reported by: rkferguson Owned by: jburel
Priority: major Milestone: 5.0.0-rc1
Component: Insight Version: 4.4.9
Keywords: n.a. Cc: m.porter@…
Resources: n.a. Referenced By: n.a.
References: n.a. Remaining Time: n.a.
Sprint: OMERO 5 Beta 2 (1)

Description

Discovered by Michael using Nightshade and 4.4.9. Confirmed and screenshots by Gus.

When you use a point ROI, the intensity value is being read off the top right hand corner - presumably the "position" of the ROI, not at the centre of the ROI i.e. the cross-hairs where you would expect it to read.

The screenshot illustrates this where the line ROI has a max intensity of 1797 in the red channel and when the point ROI is placed over the max intensity it only reads 171.

Moving the point ROI so it's top left corner is over the max intensity gives a reading of 1797.

Change History (1)

Changed 10 years ago by rkferguson

Note: See TracTickets for help on using tickets. You may also have a look at Agilo extensions to the ticket.