Task #13173 (closed)
Opened 9 years ago
Closed 8 years ago
Bug: CZI QA 17056
Reported by: | jburel | Owned by: | bramalingam |
---|---|---|---|
Priority: | major | Milestone: | B-F-5.2.0 |
Component: | Bio-Formats | Version: | OMERO-5.2.0 |
Keywords: | n.a. | Cc: | mlinkert |
Resources: | n.a. | Referenced By: | n.a. |
References: | n.a. | Remaining Time: | n.a. |
Sprint: | n.a. |
Description
https://www.openmicroscopy.org/community/viewtopic.php?f=5&t=8014
- Open the files in Zen, the images are different
- Select the directory QA 17056, only 2 images are found by importCandidates. The folder contains 9 images.
- The problem is due to the name of the files i.e. WT 0 (2).czi
- I replace the space by underscore locally
- The files are correctly detected and the images are then correctly imported.
Change History (2)
comment:1 Changed 8 years ago by bramalingam
comment:2 Changed 8 years ago by sbesson
- Resolution set to fixed
- Status changed from new to closed
Note: See
TracTickets for help on using
tickets.
You may also have a look at Agilo extensions to the ticket.
https://github.com/openmicroscopy/bioformats/pull/2323