Task #13196 (closed)
Opened 8 years ago
Closed 6 years ago
Imspector .msr FLIM t dimension seems to be interpreted as x
Reported by: | mlinkert | Owned by: | |
---|---|---|---|
Priority: | major | Milestone: | Unscheduled |
Component: | from QA | Version: | Bio-Formats-5.1.8 |
Keywords: | OBFReader, external, our-fault | Cc: | Tom.Robinson@…, i.munro@… |
Resources: | n.a. | Referenced By: | n.a. |
References: | n.a. | Remaining Time: | 0.0d |
Sprint: | n.a. |
Description (last modified by imunro-x)
@mpikg.mpg.de report that this .msr file QA 17111 cannot be opened by FLIMfit.
They are expecting 43 images each of 256x256 pixels with 266 time bins (not 256 time bins).
Bio-Formats sees 256 z-planes of 256 x 266 pixels in each image and fails to create a ModuloAlong? annotation.
Change History (6)
comment:1 Changed 8 years ago by imunro-x
- Description modified (diff)
comment:2 Changed 8 years ago by sbesson
- Keywords OBFReader added; ImspectorReader removed
comment:3 Changed 8 years ago by imunro-x
Looking through this file with a text editor the string "SPCM" occurs 73 times frequently as "SPCM Is Active". If this refers to http://www.becker-hickl.com/software/spcm.htm then this would definitely be consistent with this file being FLIM data.
comment:4 Changed 8 years ago by imunro-x
- Cc Tom.Robinson@… added
comment:5 Changed 8 years ago by imunro-x
Tom (now cc'd on the ticket has sent:
"We use a STED system from Abberior in Gottingen which has B&H TCSPC card for FLIM (which is what I am using it for).
the machine comes with this version:
http://abberior-instruments.com/products/imspectorc-software/3rd-party-compatibility/
comment:6 Changed 6 years ago by mlinkert
- Resolution set to fixed
- Status changed from new to closed
After investigation, the .msr file is not a Imspector file?(https://www.openmicroscopy.org/site/support/bio-formats5.1/formats/lavision-imspector.html) but rather an OBF file?(https://www.openmicroscopy.org/site/support/bio-formats5.1/formats/imspector-obf.html). It is unclear at this point whether the data is supposed to be containing FLIM data or not and more information is needed.