Task #2422 (closed)
Opened 9 years ago
Closed 9 years ago
Fluoview metadata
| Reported by: | jburel | Owned by: | mlinkert |
|---|---|---|---|
| Priority: | major | Milestone: | OMERO-Beta4.2 |
| Component: | Import | Version: | n.a. |
| Keywords: | n.a. | Cc: | |
| Resources: | n.a. | Referenced By: | n.a. |
| References: | n.a. | Remaining Time: | n.a. |
| Sprint: | 2010-05-27 (10) |
Description (last modified by jburel)
file is DBTRG 2x 1a1.tif
- Original Metadata: junk in the one of the header: it should be "series" or something like that but we have Ol?wLwO
- FITC: Detector Voltage was 402 now 436
- Transmitted: no Detector Voltage now 402
This means that wrong value was set. Need to investigate
Attachments (1)
Change History (8)
comment:1 Changed 9 years ago by jburel
- Description modified (diff)
- Summary changed from Fluoview Original metadata to Fluoview metadata
comment:2 Changed 9 years ago by jburel
- Description modified (diff)
comment:3 Changed 9 years ago by jburel
- Description modified (diff)
- Priority changed from minor to major
comment:4 Changed 9 years ago by cxallan
- Owner changed from cxallan to mlinkert-x
comment:5 Changed 9 years ago by cxallan
- Sprint changed from 2010-05-13 (9) to 2010-05-27 (10)
comment:6 Changed 9 years ago by mlinkert
Garbage data in original metadata should be fixed in bio-formats:r6404.
Detector Voltages have been fixed in bio-formats:r6405, but note that they will be different from the values in 4.1. I opened this file in the Olympus Fluoview software (skyking:/data/software/olympus/Viewer/), and found that the correct values were:
FITC: no voltage
Transmitted: 436
See attached screenshot.
Changed 9 years ago by mlinkert-x
comment:7 Changed 9 years ago by cxallan
- Resolution set to fixed
- Status changed from new to closed
Fixed in Trunk, verified with bio-formats:r6414.
Checked with bio-formats:r6389; still a problem. Verified in ImageJ with loci_tools.jar.