Notice: In order to edit this ticket you need to be either: a Product Owner, The owner or the reporter of the ticket, or, in case of a Task not yet assigned, a team_member"

Task #8094 (closed)

Opened 7 years ago

Closed 7 years ago

Last modified 7 years ago

BUG:Scale bar split view

Reported by: omero-qa Owned by: jburel
Priority: minor Milestone: OMERO-4.4
Component: from QA Version: n.a.
Keywords: n.a. Cc: a.vonderheit@…
Resources: n.a. Referenced By: n.a.
References: n.a. Remaining Time: n.a.
Sprint: 2012-03-27 (11)

Description

http://qa.openmicroscopy.org.uk/qa/feedback/4218/

Comment: Dear OME-Team,
we are testing OMERO for using it in our Core Facility.
I just stumble over a small error.
The scale bar changes when I use the Split tab, the scale bar is much smaller as in the Image or the Projection tab.
The Images (z-stack) came from a Leica lif-file.
Best regrads
Andreas


    

Attachments (2)

OME-Error01.png (613.9 KB) - added by jburel 7 years ago.
OME-Error02.png (1.4 MB) - added by jburel 7 years ago.

Change History (5)

Changed 7 years ago by jburel

Changed 7 years ago by jburel

comment:1 Changed 7 years ago by jburel

  • Milestone changed from Unscheduled to OMERO-Beta4.4
  • Sprint set to 2012-03-27 (11)
  • Status changed from new to accepted

comment:2 Changed 7 years ago by jburel

  • Resolution set to fixed
  • Status changed from accepted to closed

comment:3 Changed 7 years ago by jburel <j.burel@…>

(In [7fa1f50a919568f69c07774ceb5edf0d2b4a9985/ome.git] on branch develop) Fix scale bar size (see #8094)

The value was wrongly multiplied twice by the grid ratio.