Bug: units are not correctly parsed from Gatan .dm3 files
| Reported by: | mlinkert | Owned by: | mlinkert |
|---|---|---|---|
| Priority: | major | Milestone: | 5.0.2 |
| Component: | Bio-Formats | Version: | 5.0.0 |
| Keywords: | n.a. | Cc: | ngdias@…, levachevais@… |
| Resources: | n.a. | Referenced By: | n.a. |
| References: | n.a. | Remaining Time: | n.a. |
| Sprint: | n.a. |
Description
See http://lists.openmicroscopy.org.uk/pipermail/ome-users/2014-April/004319.html and QA 8007. The issue is that the reader uses the last stored units value when converting the scale values to micrometers, when it should use the units value stored immediately after the scale value(s) being used.
Change History (3)
comment:1 Changed 5 years ago by bramalingam
comment:2 Changed 5 years ago by bramalingam
- Cc levachevais@… added
comment:3 Changed 5 years ago by mlinkert
- Resolution set to fixed
- Status changed from new to closed
Fixed with https://github.com/openmicroscopy/bioformats/pull/1079; a new build is not yet available, but this should be included in 5.0.2.